← Vulnerability feed

Vulnerability record · CVE-2017-12725 · published 15 February 2018

CVE-2017-12725: Smiths-medical medfusion 4000 wireless syringe infusion pump hard-coded credentials vulnerability

SSmiths Medical · Medfusion 4000 Wireless Syringe Infusion Pump

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

5.6 CVSS 3.0 Medium EPSS 0.83% · top 44.0% CWE-798 · Hard-coded credentials
5.6CVSS 3.0 base score, v2 6.8
0.83%EPSS exploitation probability, 30 days
NoNot in CISA KEV
1Affected product versions listed by NVD
4References
17 Jun 2026Last modified by NVD

Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L

Affected products

1 vulnerable configurations from NVD's CPE data, grouped by vendor and product.

References

LinkTags
http://www.securityfocus.com/bid/100665 Third Party AdvisoryVDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A Third Party AdvisoryUS Government Resource
http://www.securityfocus.com/bid/100665 Third Party AdvisoryVDB Entry
https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A Third Party AdvisoryUS Government Resource

Track CVE-2017-12725 inside VULONE

Watch it alongside the ransomware crews, C2 infrastructure and forum chatter that reference it, query it through the API and pull it into your SIEM over TAXII.

Start free Open in platform

Related vulnerabilities

Same products first, then exploited flaws of the same weakness class.

8.1CVE-2017-12718Smiths-medical medfusion 4000 wireless syringe infusion pump classic buffer overflow vulnerabilityA Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-pa…EPSS 13%8.1CVE-2017-12720Smiths-medical medfusion 4000 wireless syringe infusion pump missing authentication for critical function vulnerabilityAn Improper Access Control issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The FTP s…EPSS 1.9%8.1CVE-2017-12724Smiths-medical medfusion 4000 wireless syringe infusion pump hard-coded credentials vulnerabilityA Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The …EPSS 1.3%7.3CVE-2017-12726Smiths-medical medfusion 4000 wireless syringe infusion pump hard-coded credentials vulnerabilityA Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet …EPSS 1.1%5.9CVE-2017-12721Smiths-medical medfusion 4000 wireless syringe infusion pump improper certificate validation vulnerabilityAn Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. T…EPSS 0.67%5.3CVE-2017-12722Smiths-medical medfusion 4000 wireless syringe infusion pump out-of-bounds read vulnerabilityAn Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party …EPSS 2.4%3.7CVE-2017-12723Smiths-medical medfusion 4000 wireless syringe infusion pump information exposure vulnerabilityA Password in Configuration File issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The…EPSS 0.98%10.0CVE-2026-22769Dell RecoverPoint for Virtual Machines hardcoded credential flawDell RecoverPoint for Virtual Machines versions prior to 6.0.3.1 HF1 contain a hardcoded credential vulnerability (CWE-798). An unauthenticated remot…KEVEPSS 13%analysed

Source: NIST National Vulnerability Database (record CVE-2017-12725), CISA KEV, FIRST EPSS (scores of 2026-09-28). This page is refreshed as NVD updates the record.