Vulnerability record · CVE-2015-3459 · published 29 April 2015
CVE-2015-3459: Hospira lifecare pcainfusion firmware permissions and access controls vulnerability
Hospira · Lifecare Pcainfusion Firmware
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
Description
The communication module on the Hospira LifeCare PCA Infusion System before 7.0 does not require authentication for root TELNET sessions, which allows remote attackers to modify the pump configuration via unspecified commands.
AV:N/AC:L/Au:N/C:C/I:C/A:C